§ 221.37a. Activation of fluoroscopic tube.
X-ray production in the fluoroscopic mode shall be controlled by a device which requires continuous pressure by the fluoroscopist for the entire time of the exposure (dead-man switch). When recording serial fluoroscopic images, the fluoroscopist shall be able to terminate X-ray exposures at any time, but means may be provided to permit completion of any single exposure of the series in process.
Authority The provisions of this § 221.37a issued under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.37a adopted October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894.
Cross References This section cited in 25 Pa. Code § 221.35a (relating to fluoroscopic X-ray systems); 25 Pa. Code § 221.43a (relating to mobile fluoroscopes); and 25 Pa. Code § 221.61 (relating to radiation therapy simulation systems).
No part of the information on this site may be reproduced for profit or sold for profit.
This material has been drawn directly from the official Pennsylvania Code full text database. Due to the limitations of HTML or differences in display capabilities of different browsers, this version may differ slightly from the official printed version.