§ 221.65. X-ray attenuation systems.
CT systems solely used to calculate attenuation coefficients or for image registration in nuclear medicine studies must meet the requirements in § § 221.202221.205 unless otherwise exempted as follows:
(1) CT systems identified in this section are exempt from § § 221.202(a) and 221.204(a)(4)(xi) (relating to equipment requirements; and performance evaluations, routine QC and surveys).
(2) Instead of § 221.204(a) (relating to performance evaluations, routine QC and surveys), the registrant shall complete a performance evaluation on the CT system following the recommendations of a QMP, the system manufacturer or a nationally-recognized organization at intervals not to exceed 14 months.
(3) Instead of § 221.204(b), checks shall be established and documented by the registrant following nationally-recognized guidelines or those recommended by the manufacturer.
Authority The provisions of this § 221.65 issued under sections 301 and 302 of the Radiation Protection Act (35 P.S. § § 7110.301 and 7110.302); and section 1920-A of the Administrative Code (71 P.S. § 510.20).
Source The provisions of this § 221.65 adopted October 26, 2018, effective January 24, 2019, 48 Pa.B. 6791.
No part of the information on this site may be reproduced for profit or sold for profit.
This material has been drawn directly from the official Pennsylvania Code full text database. Due to the limitations of HTML or differences in display capabilities of different browsers, this version may differ slightly from the official printed version.