Pennsylvania Code & Bulletin
COMMONWEALTH OF PENNSYLVANIA

• No statutes or acts will be found at this website.

The Pennsylvania Code website reflects the Pennsylvania Code changes effective through 54 Pa.B. 5598 (August 31, 2024).

25 Pa. Code § 228.64. Filters.

§ 228.64. Filters.

 (a)  A filter which is removable from the system shall be clearly identified. Documentation shall contain a description of the filter which includes a drawing showing dimensions and noting materials of construction.

 (b)  For new equipment which utilizes a system of wedge filters, interchangeable field flattening filters or interchangeable beam scattering filters the following apply:

   (1)  Irradiation may not be possible until a selection of a filter has been made at the control panel.

   (2)  An interlock system shall be provided to prevent irradiation if the filter selected is not in the correct position.

   (3)  An interlock shall be provided to prevent irradiation if a filter selection operation carried out in the treatment room does not agree with the filter selection operation carried out at the control panel.

Source

   The provisions of this §  228.64 adopted October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894.



No part of the information on this site may be reproduced for profit or sold for profit.


This material has been drawn directly from the official Pennsylvania Code full text database. Due to the limitations of HTML or differences in display capabilities of different browsers, this version may differ slightly from the official printed version.