Pennsylvania Code & Bulletin
COMMONWEALTH OF PENNSYLVANIA

• No statutes or acts will be found at this website.

The Pennsylvania Code website reflects the Pennsylvania Code changes effective through 54 Pa.B. 6234 (September 28, 2024).

Pennsylvania Code



Subchapter D. INSTRUMENTATION


Sec.


157.31.    Types of measurement systems which may be used.
157.32.    Calibration of measurement systems.
157.33.    Windscreens.

§ 157.31. Types of measurement systems which may be used.

 The sound level measurement system shall meet or exceed the requirements of American National Standard Specifications for Sound Level Meters, ANSI S1.4—1971, approved April 27, 1971, issued by the American National Standards Institute, throughout the applicable frequency range for either:

   (1)  A Type 1 sound level meter.

   (2)  A Type 2 sound level meter.

   (3)  A Type S sound level meter which has the following:

     (i)   A-weighting frequence response.

     (ii)   Fast dynamic characteristics of its indicating instrument.

     (iii)   A relative response level tolerance consistent with those of either a Type 1 or Type 2 sound level meter, as specified in Section 3.2 of ANSI S1.4—1971.

Cross References

   This section cited in 67 Pa. Code §  157.42 (relating to ambient conditions); and 67 Pa. Code §  157.43 (relating to location and operation of sound level measurement system).

§ 157.32. Calibration of measurement systems.

 (a)  Calibration. Requirements for calibration shall be as follows:

   (1)  The sound level measurement system shall be calibrated and appropriately adjusted at one or more frequencies in the range from 250 to 1,000 hertz at the beginning of each series of measurements and at intervals of 5—15 minutes thereafter until it has been established that the sound level measurement system has not drifted from its calibrated level. Once this fact has been established, calibrations may be made at intervals of once every hour. A significant drift shall be considered to have occurred if a 0.3 decibel or more excursion is noted from the systems predetermined reference calibration level. In the case of systems using displays with whole decibel increments, the operator may usually judge when the 0.3 decibel drift has been met or exceeded.

   (2)  The sound level measurement system shall be checked periodically by its manufacturer, a representative of its manufacturer, or a person of equivalent special competence to verify that its accuracy meets the design criteria of the manufacturer.

 (b)  Acoustical calibrator. An acoustical calibrator of the microphone coupler type design for the sound level measurement system in use shall be used to calibrate the sound level measurement system in accordance with subsection (a) of this section. The calibrator shall meet or exceed the accuracy requirements specified in section 5.4.1 of the American National Standards Institute Standard Methods for Measurements of Sound Pressure Levels, ANSI §  1.13—1971, for calibrators for field type measurements.

§ 157.33. Windscreens.

 A properly installed windscreen, of the type recommended by the manufacturer of the sound level measurement system, shall be used during the time that noise emission measurements are being taken.



No part of the information on this site may be reproduced for profit or sold for profit.


This material has been drawn directly from the official Pennsylvania Code full text database. Due to the limitations of HTML or differences in display capabilities of different browsers, this version may differ slightly from the official printed version.